Chapter 4: Performance Features and Safety Precautions

Product Service 1-800-522-7658

27

• Performance tested to maintain oxygen levels as low as 0.2 ppm and

moisture levels as low as 1 ppm.

• Oxygen permeation <0.16 ppm/min per ISO 10648-2. Exceeds Class

1 conditions at <1.67 ppm/min.

• Illumination exceeds 80 foot candles.

• Doublewide models include 4 glove ports with main chamber volume

of 34 cu. ft.

3. Basic Manual Control Models include:

• Main chamber pressure analog gauge with readings from -5 to +5

inches of water gauge.

• Transfer chamber pressure analog gauge with readings from 0 to -29.5

inches of mercury.

• Switch to control exterior outlet for vacuum pump connection.

4. Auto Pressure Controller Model include:

• Automatic control of the main chamber pressure (-5 to +5 inches of

water gauge), main chamber evacuation and filling cycles (up to 499
cycles), and transfer chamber evacuation and filling cycles (up to 199
cycles).

• Regulation of vacuum/pressure between -5 and +5 inches of water

gauge in the main chamber and between 0 and -29.5 inches of
mercury in the transfer chamber.

• Mutable, audible alarm that sounds when pressure/vacuum level is

outside set limits.

• LCD display of main chamber and transfer chamber pressure in

inches of water, inches of mercury and mBar.

• Three-position foot pedal (increase gas/fill, off, evacuation/decrease

gas).

• C.A. COM (6-Pin) port for shared vacuum pump use with the optional

AtmosPure™ Re-Gen Air Purifier.


All models feature:

5. Large viewing area, 36" w x 27.6" high, provides maximum visibility.

Removable viewing window is 3/8" thick, laminated safety glass. Double
wide models have two viewing areas.

6. Pair of neoprene gloves, 0.015" thick, 30" long, size 9 ¾. Doublewide

models include two pair. See Replacement Parts for all glove types and
sizes.

7. Space-saving inner and outer transfer chamber doors that pivot upward,

counterbalanced and equipped with quick-latches.

8. Four manual valves – two on the main chamber and two on the transfer

chamber – for evacuating, filling and manual control. With 3/8"
compression fittings and 3/8" hose barbs.

9. Factory-installed, rear-mounted, low profile stainless steel pressure relief

bubbler at ±6 inches of water gauge.

10. Two rear-mounted inlet and outlet 3/4" OD compression fittings for easy

connection to optional AtmosPure Re-Gen Air Purifier or other accessory
drying train.