Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

39

Use the “MODE” key in the upper left hand corner adjacent to the icon to return to
the “Glove Box Pressure Control” screen to use the “Select” mode.

Transfer Chamber Evacuation (Purge) and Fill –
Operation Mode 3

The Transfer Chamber Purge/Fill mode controls the pressure changes inside the
transfer chamber. Pressure in the transfer chamber will be displayed inside the
transfer chamber icon on the display screen. This operation mode controls the gas
inlet and gas outlet solenoid valves connected to the transfer chamber.

























The user may select up to 199 “Purge/Fill” cycles by adjusting the cycles increase
and decrease arrow keys. The user can select the level of pressure inside the transfer
chamber during each purge between 0 and -29.5 inches of Mercury/Hg [0 -999
mBars] by adjusting the lower limit keys. Once the lower pressure limit and number
of cycles have been selected, press the key adjacent to “Start,” and the automated
purge and fill process will begin and continue for the number of cycles selected.
Each cycle will be displayed during the process. To cancel the automated purge and
fill process, press the key adjacent to “Stop” and the process will be stopped.

Manual

Purge

Manual

Fill

Start/Stop

Setup

Increase
Cycles

Decrease
Cycles

Mode Key

Mute/Alarm

Increase

Lower Limit

Decrease

Lower Limit