Chapter 1: Introduction
Product Service 1-800-522-7658
2
Glove
Gas
Port
Permeant
Covers
Protector (FRP) Protector (stainless)
Attainable
O
2
Installed
0.4
0.2
Purity
H
2
O
Installed
2
1
(ppm)
O
2
None
0.9
0.8
H
2
O
None
3
2
Oxygen Permeation
O
2
None
0.30
0.16
at -4" w.g. per
ISO 10648-2
Test Method (ppm/min)
ISO 10648-2
O
2
None
<1.67
<1.67
Acceptance Level
for Class 1 Controlled
Atmosphere Glove Box
(ppm/min)
Usable
Permeation
O
2
None
0.16
0.14
with pressure from
H
2
O
None
0.17
0.10
(-1" w.g. to +1" w.g.)
(ppm/min)
Glove Box Type
The Protector Glove Boxes offer many unique features to enhance
performance. To take full advantage of them, you must read and understand
this manual and keep it handy for future reference. If you are unfamiliar with
how the Glove Boxes operate, please review Chapter 4: Performance
Features and Safety Precautions before you begin working in the glove box.
Even if you are an experienced user, please review Chapter 5: Using Your
Glove Box so you can use the glove box efficiently.