Refer to the following table if your glove box fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Unit not plugged into outlet.

Plug unit into appropriate electrical service.

Circuit breaker(s) tripped.

Reset or replace circuit breaker.

Lamp not installed properly.

Inspect lamp installation.

Lamp wiring disconnected.

Inspect lamp wiring.

Defective lamp.

Replace lamp.

Light switch is defective.

Replace light switch.

Lights will not operate.

Defective electronic ballast.

Replace ballast.

Unit not plugged in.

Plug unit into electrical service.

Circuit breaker(s) tripped.

Reset or replace circuit breaker.

Receptacle switch is
defective.

Replace receptacle switch.

No power to electrical
receptacle.

Receptacle is defective.

Replace receptacle.

Plumbing is installed
incorrectly.

Check plumbing and insure VAC is “out”
to vacuum pump and GAS is “in” from
inert gas supply.

Auto Pressure Controller
Solenoid Valves will not
operate.

Solenoid Valve is defective.

Replace solenoid valve.

Unit not plugged into outlet.
Circuit breaker(s) tripped.

Plug unit into appropriate electrical service.
Reset or replace circuit breaker.

Auto Pressure Controller
will not operate.

LCD Display is defective.

Replace circuit board or power supply
circuit board.