Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

40

Setup Menu – Operation Mode 4

Pressing the key opposite the word “Setup,” allows access to the setup operation
mode. In the setup mode, the user may select their preferred units of pressure that are
displayed, restore settings back to the default state, or calibrate the pressure sensors.
The keys adjacent to the up and down arrows on the lower right of the display are
used to select one of the three options. After toggling up or down, push the NEXT
key to select the option.
























Option 1
Units of Measure: If this option is selected, the key adjacent to “Next” is used
to select pressure units of inches of water, inches of mercury or mBar.

Helpful conversions are as follows:

1 mBar = .39 inches of water column
1 mbar = .0292 inches of mercury
1 inch of water = 2.52 mBar = .0737 inches of mercury
1 inch of mercury = 34.25 mBar

All pressures are measured with respect to atmosphere pressure.