Main Modes of Operation
The Labconco Glove Box Pressure Controller display has four main screens
that provide access to the 4 modes of operation listed below.
1.) Glove Box Pressure Control Mode
2.) Glove Box Purge and Fill Mode
3.) Transfer Chamber Purge and Fill Mode
4.) Setup Mode
Press the MODE key at the upper-left hand corner of the display screen to
switch between each of these four operation modes. Please be aware that if
one of the glove box modes is running/started, other glove box operation
modes cannot be used. For example, if the Glove Box Pressure Control mode
has started (using the start key), the menu page for the Glove Box Purge/Fill
mode cannot be used – this prevents operational control conflicts from
occurring at the same time.
Manual
Purge
Manual
Fill
Start/Stop
Setup
Increase
Upper Limit
Decrease
Upper Limit
Mode Key
Mute/Alarm
Increase
Lower Limit
Decrease
Lower Limit