Chapter 2: Prerequisites

Product Service 1-800-522-7658

6

Labconco does not recommend hard duct connections with remote
blowers due to the air volume balance requirements with such systems.
Hard duct connections should only be used when the glove box impeller
is disconnected and only a single roof-mounted blower is used. Note that
reliance on an external exhaust blower does not provide airflow control
at the glove box.

If a positive internal pressure clean air ISO Class 3 condition (ISO 14644-1)
is desired, the motorized impeller can be rotated to work in positive pressure
mode (see Figure 3-2). See Chapter 7 for the Positive Pressure Conversion
Kit. See the following Negative and Positive Pressure Particle Test data in
the charts below. Both charts show the fast dilution rates of the glove boxes.

Protector HEPA-Filtered Glove Box Particle Testing under Negative Pressure

0

143

144

123

123

107

103

103

100

0

400

38

6

8

7

5

9

8

2

5

0

50

100

150

200

250

300

350

400

450

500

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

Time (Minutes)

N

u

m

b

er

o

f Partic

le

s

(P

er C

u

b

ic M

e

te

r Per M

inu

te)

> 0.3 Micron

> 0.5 Micron

Airflow Testing Performed at 42 CFM

with Glove Box Under Negative Pressure.

NOTE: Ambient Air Baseline Particle Count at 1 Minute
132155 Particles measuring 0.3 Micron detected within 1 Minute.
6499 Particles measuring 0.5 Micron detected within 1 Minute.