Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

30

Main Setup Screens





































RESTORE DEFAULTS

All User Data will be reset
except calibration.
UP to Select
MENU to Escape

CALIBRATION LOCK

▲▼ to Choose, ENTER
to Select, MENU to Exit

0

SETUP (cont.)

1. Units of Measure
2. Security Lock
3. Calibration
4. Restore Defaults
5. Diagnostic

DIAGNOSTIC TEST

▲▼ to Choose, ENTER
to Select, MENU to Exit
f 1. Keypad
2. Airflow Sensor
3. Pressure Sensor

Certifier’s Use Only!

234 code to access

calibration mode.

KEY PAD TEST

All Keys Are Off.

AIRFLOW SENSOR

X1:

258/258

X5:

218/218

X5 CAL:

135/218


Airflow:

79/ 80 CFM

PRESSURE SENSOR

Press:
X1: 286
CAL: 258/286

Pressure: -1.99/1.97 IOW

Enter

Enter

Enter

Enter

Enter

Enter

See “Diagnostics Program Screens”
for more info.