Main setup screens, Chapter 5: using your glove box – Labconco Protector Filtered Glove Box User Manual
Page 34
![background image](/manuals/738717/34/background.png)
Chapter 5: Using Your Glove Box
Product Service 1-800-522-7658
30
Main Setup Screens
RESTORE DEFAULTS
All User Data will be reset
except calibration.
UP to Select
MENU to Escape
CALIBRATION LOCK
▲▼ to Choose, ENTER
to Select, MENU to Exit
0
SETUP (cont.)
1. Units of Measure
2. Security Lock
3. Calibration
4. Restore Defaults
5. Diagnostic
DIAGNOSTIC TEST
▲▼ to Choose, ENTER
to Select, MENU to Exit
f 1. Keypad
2. Airflow Sensor
3. Pressure Sensor
Certifier’s Use Only!
234 code to access
calibration mode.
KEY PAD TEST
All Keys Are Off.
AIRFLOW SENSOR
X1:
258/258
X5:
218/218
X5 CAL:
135/218
Airflow:
79/ 80 CFM
PRESSURE SENSOR
Press:
X1: 286
CAL: 258/286
Pressure: -1.99/1.97 IOW
Enter
Enter
Enter
Enter
Enter
Enter
See “Diagnostics Program Screens”
for more info.