Diagnostic program screens – Labconco Protector Filtered Glove Box User Manual

Page 36

Advertising
background image

Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

32

Diagnostic Program Screens
























DIAGNOSTIC TEST

▲▼ to Choose, ENTER
to Select, MENU to Exit
f 1. Keypad
2. Airflow Sensor
3. Pressure Sensor

AIRLFOW SENSOR

X1:

261/XXX

X5:

212/XXX

X5 CAL:

134/212

Airflow:

70/70 CFM

KEY PAD TEST

All Keys Are Off.

Airflow Sensor Diagnostic Notes:
X5 CAL example represents a pressure of
134 with the blower off and 212 with the
blower speed to produce 70 CFM for this
calibration example. XXX, 134, and 212
are calibrated values for this example.

Pressure Sensor Diagnostic Notes:
X1 example indicates raw voltage
level of pressure sensor at present
conditions. CAL example represents
a raw pressure of 258 with the doors
in the glove box open and no
pressure inside the glove box and a
raw pressure of 286 when the
negative high pressure level was
calibrated at -1.97 inches of H

2

O

(IOW). Press example shows a
current pressure of -1.99 IOW and
the unit’s high calibration point of
1.97 IOW.

PRESSURE SENSOR

Press:
X1: 286
CAL: 258/286

Pressure: -1.99/1.97 IOW