Chapter 5: Using Your Glove Box

Product Service 1-800-522-7658

31

Calibration Program Screens











































Enter code for

certification use only and

then press

ENTER

key

CALIBRATION LOCK

?? to Choose, ENTER

to Select, MENU to Exit

0

AIRFLOW SENSOR ZERO

Please Wait:

150

AIRFLOW SENSOR LEVEL ADJUST

Set to Known Airflow

?? to Choose, ENTER to Select,

MENU to Exit

Airflow:

70

CFM

CALIBRATION

??

to Choose,

ENTER to Select,

MENU to Exit

f

1. Airflow Sensor

2. Airflow Alarm

3. Pressure Sensor

4. Pressure Alarm

5. Box Size

6. Box Type

ALARM LEVEL

?? to Choose, ENTER

to Select, MENU to Exit

Alarm Level: 10 CFM

Alarm Level is best set at low levels such

as 10 CFM below oper

ating speed.

PRESSURE SENSOR

1. Zero Adjust

2. Level Adjust

PRESSURE ALARM

?? to Choose, ENTER to

Select, MENU to Exit

Alarm Level: X.XX IOW

PRESSURE SENSOR

ZERO ADJUST

Open all doors

Enter when done

PRESSURE SENSOR

ZERO ADJUST

Wait 150

Seconds

PRESSURE SENSOR

LEVEL ADJUST

Set known Pressure

?? to Choose, ENTER to

Select, MENU to Exit

Pressure: X.XX IOW

AIRFLOW ALARM

1. Alarm Level

2. Alarm Delay

BOX SIZE

1. Single

2. Double

AIRFLOW SENSOR

1. Zero Adjust

2. Level Adjust

ALARM DELAY

?? to Choose, ENTER to
Select, MENU to Exit

Alarm Level: xxx Seconds

PRESSURE ALARM

1. Alarm Level

2. Alarm Delay

ALARM DELAY

?? to Choose, ENTER to

Select, MENU to Exit

Alarm Level: xxx Seconds

Pressure Sensor Auto-Zero Note: Every time blower is shut off, wait 100 seconds and thentake zero data for 30 seconds.
Auto-

Zero occurs while the blower is off for the entire time span. The “span” or difference between zero airflow and 70

CFM

is saved from “Pressure Sensor” calibration in step 2. Also, flash icon on LCD to shut Auto-Zero.

AIRFLOW SENSOR

AUTO-ZERO

1. Disable

2. Enable

AIRFLOW SENSOR ZERO

Please Wait:

150

Enable

BOX TYPE

1. Positive Pressure

2. Negative Pressure

PRESSURE SENSOR

LEVEL ADJUST

Wait 150 Seconds