Motorized Impeller Replacement

The motorized impeller must be replaced as a complete unit. See Appendix A
for replacement parts diagram. Consult your Safety Officer for proper
decontamination or sealing of tubing before replacing the blower.

1. Unplug the motorized impeller from the electrical outlet. Refer to

Appendix A for replacement parts.

2. Remove the hose connecting the exhaust filter to the motorized impeller.
3. Remove the motorized impeller and support braces.
4. Label and remove the tubing as shown in Figure 6-3.
5. Reassemble the new motorized impeller by reversing the steps above.

WARNING: High-speed blower. Never operate impeller
with housing off.

Figure 6-3

Motorized Impeller & Tubing

!

Rear Port Blower
Static Pressure S.P.

Front Port Blower
Total Pressure T.P.

Silicone Larger
I.D. Tubing

Exhaust Filter

Housing

Main Chamber
Static Pressure S.P.

Connection

Hose

Motorized
Impeller
Assembly

T.P.

S.P.