Chapter 3: Getting Started

Product Service 1-800-522-7658

21

• Static pressure measured inside the glove box varied from negative .12

inches of water column at 15-CFM to negative 1.28 inches of water at 84-
CFM. The total system static pressure as measured between the exhaust
filter and the blower varied from negative 0.31 inches of water column at
15-CFM to negative 2.48 inches of water at 84-CFM.

• Displacement (vibration level) measured on the work surface is

negligible at minimum through maximum airflow volume settings. The
measured level of displacement ranged from 0.07 x 10

-5

meters to 0.38 x

10

-5

meters.

• Balance stability test results were excellent with no detectable instability

of an analytical balance to five decimal places and a top loader to three
decimal places. Test results were confirmed with the glove box operating
at various airflow volume settings and with various sample weights.