Chapter 2: prerequisites – Labconco Protector Filtered Glove Box User Manual

Page 12

Advertising
background image

Chapter 2: Prerequisites

Product Service 1-800-522-7658

8

Additionally, the motorized impeller can also be used in a full recirculation
mode providing the operator with some control of the atmosphere chemistry
within the glove box. See Chapter 7 for the Recirculation Conversion Kit.

The airflow charts below represents typical operational data including air
changes per minute, exhaust volume, noise levels, main chamber static
pressures, and total system static pressures measured at the blower intake
(suction side). Exhaust volumes are reduced by 17% for products operating
at 230V. Exhaust volumes are reduced by 22% for ULPA filters. Exhaust
volumes are reduced by 17% for products operating on 100V, 60 Hz and
reduced 34% for 100V, 50 Hz products.

% Blower Speed

Control Setting

Air Changes

per Minute

(ACM)

HEPA Exhaust

Volume

(CFM)

Noise Pressure

dB (A)

Static Pressure

Inside Glove Box

(inches of water)

Total System

Static Pressure

(inches of water)

Min. 18%

35%
55%
75%

Max. 100%

Dial OFF

0.9
1.8
2.8
3.8
5.0

0

15
30
48
65
85

0

52
54
59
63
68
37

0.12
0.30
0.54
0.87
1.28

0

0.31
0.62
1.14
1.65
2.48

0

PROTECTOR HEPA-FILTERED GLOVE BOX

Static Pressure Test Results

0.12

0.19

0.27

0.36

0.52

0.68

0.87

1.02

1.28

0.31

0.40

0.57

0.72

1.12

1.39

1.65

1.94

2.48

0

0.5

1

1.5

2

2.5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

Airflow Volume

(Cubic Feet per Minute)

Nega

ti

ve Sta

ti

c

Pre

ssure

(In

che

s of

Wat

e

r Colum

n

)

Static Pressure Inside Glove Box

Total Static Pressure at Blower
Intake (Suction Side)

Static Pressure test results on the Protector HEPA-Filtered Glove Box

Note: Test results are approximately the same for positive pressure application