Static Pressure Test Results

0.30

0.51

0.72

0.83

1.00

0.52

0.89

1.23

1.41

1.67

0

0.5

1

1.5

2

2.5

20

25

30

35

40

45

50

55

60

Airflow Volume

(Cubic Feet per Minute)

Nega

tiv

e St

at

ic

Pressure

(I

nche

s of

Water

Col

umn)

Static Pressure Inside Glove Box

Total Static Pressure at Blower
Intake (Suction Side)

Static Pressure test results on the Protector ULPA-Filtered Glove Box

Note: Test results are approximately the same for positive pressure application

PROTECTOR ULPA-FILTERED GLOVE BOX (230V)