13

SETUP

Inserting and Connecting the

Batteries

Open the torso as described in steps 1 - 4, Opening the Torso.
To remove the batteries, follow the same procedure in reverse.

1 Release the battery clamp by unhooking the clips on either side.

2 Insert both batteries into the battery tray.

3 Snap the battery clamp back into place over the batteries.

4 Connect the corresponding battery cables from the batteries to

the torso.

After connecting the batteries, connect the Patient Simulator to the
external power supply (12V to 24V) while turned OFF.

The batteries will charge if the Patient Simulator is ON and
connected to an external power supply in the range of (20V - 24V).

Battery - Cable and Tube Descriptions

Name/Label

Tube/Cable Color

Connector Description

Battery 1

Black harness cable

Black rectangular
connector, 6 lead

Battery 2

Black harness cable

Black rectangular
connector, 6 lead