7 Reverse this process to insert a new lung.

Note: Ensure that the compliance bands intersect between the
two folds of the lung.

Replacing Simulator Skins

The simulator skins may need to be replaced if they become torn,
perforated, or stained.

1 Unzip and unroll the skin.

2 Applying powder to the inside of arm, torso and leg skins will

help reduce friction and make replacing the skins easier.

To prevent the zippers from separating from the skin, make sure
to properly position the skin halves and hold them together while
zipping them closed.

Replacing Blood System Filter

If reduced blood flow is experienced, the filter may be clogged and
needs replacement.

Never run the Patient Simulator without a filter.

Bloo

d R

igh

t L

eg

1 Turn off the Patient Simulator.

2 Remove genitalia with the catheterization assembly for easy

access.

3 Disconnect the filter from the right leg and pelvis blood tubes

and remove it.

4 Connect a new filter by reversing the steps above.

Servicing the Patient Simulator

A full service, including cleaning of the base board, should be
performed at regular intervals.

Caution: All servicing must be performed by qualified service

personnel.

Always perform a service:

− If liquids have been spilled in the Patient Simulator

− After use in dusty environments.

Caution: Do not use cables or connectors showing visible damage.

41

MAINTENANCE