Air/co, Panel, Blood and fluid system – Laerdal SimMan 3G Trauma User Manual

Page 17: Setup, Panel has 2 connection ports, Connecting external air and co, Supply, Simman 3g trauma right leg fill panel, Fill internal fluid reservoir, Empty internal fluid reservoir

Advertising
background image

17

SETUP

Air/CO

2

Panel

The Air/CO

2

panel is located on the left side of the torso. To access

the panel, lift up the simulator skin flap and remove the protective
covering. Connect external Air/CO

2

.

Air/CO

2

panel has 2 connection ports

Air / CO

2

Inlet

Blood

Outlet

Connecting External Air and CO

2

Supply

An internal compressor is located in the right leg of the Patient
Simulator. It is recommended to use an external source of
compressed air whenever the Patient Simulator is stationary over
extended periods of use.

Connect CO

2

only if the Patient Simulator is required to exhale

CO

2

with each ventilation. Exhaled CO

2

can be detected with a real

capnographic device. The Patient Simulator will only exhale CO

2

when a capnograph is registered as being connected to the system.

1 Connect a suitable CO

2

source to a Laerdal external

compressor or regulator panel.

2 Connect a Laerdal double-lumen Air/CO

2

tube from the

external compressor or regulator panel to the Air/CO

2

inlet on

the panel.

For more information on external compressors and regulator
panels compatible with SimMan 3G Trauma, contact your local
Laerdal representative.

Blood and Fluid System

The Patient Simulator has two internal reservoirs, one for blood and
one for fluids/secretions. SimMan 3G Trauma is also supplied with
two fill units - one Blood Fill Unit for blood and one Fluid Fill Unit for
fluids /secretions.

SimMan 3G Trauma Right Leg Fill Panel

The right leg fill panel is located at the top of the right leg near the
pelvis. The fill panel contains connectors for filling the blood and fluid
reservoirs.

Note: Make sure the Patient Simulator power is on.

Fill Internal Fluid Reservoir

1 Roll the right leg skin down to expose the fill panel.

2 Connect fluid fill unit tubes to the fluid and air connectors

in the right leg panel.

3 Push the fill button on the panel. The button will light up and

fluid will flow into the Patient Simulator.

4 When the flow stops, disconnect the fill unit.

5 Push the fill button on the panel. The light will go out.

Note: Disconnect tubes from the Patient Simulator before pushing
the fill button. Pushing the button before disconnecting the tubes
will initiate draining of the tank.

Empty Internal Fluid Reservoir

1 Connect an empty fluid fill unit to fluid connector in

the right leg panel.

2 Fluid from the internal reservoir will drain into the bottle.

3 When the flow stops, disconnect the fluid connector.










Run Patient Simulator with External Fluid

1 Drain the internal reservoir. Follow instructions for “Empty

Internal Fluid Reservoir”.

2 After draining the internal reservoir, fill the fluid fill unit and

connect to the Patient Simulator.

3 Push the fill button on the panel. The button will light and fluid

will flow into the Patient Simulator.

4 Charge the system for 60 seconds before starting simulation.

Warning: Connecting a full fluid fill unit to a Patient Simulator with a

full internal reservoir will result in system overflow. Fluid will drain
out of the right leg. Repeatedly overflowing the system may
damage the product.

Fluid

Inlet

Air Out

Blood

Inlet

Activate

Filling

Fluid

Inlet

Air Out

Blood

Inlet

Activate

Filling

Air Out

Blood

Inlet

SimMan

Fluid Fill Unit

Cautions

and

W

arnings

Featur

es

Setup

Maintenance

Spar

e Par

ts

Tr

oubleshooting