Battery Warnings

Warning: Do not run the Patient Simulator for more than 1 minute

on a single battery.

Warning: If both batteries are removed while the simulation is

paused, the Patient Simulator will shut down and simulation data
will be lost.

Warning: Dispose of batteries in accordance with local regulations.

Warning: The external battery charger is for indoor use only.

Warning: The batteries should only be charged in temperatures

ranging from 0 °C - 40°C (32 °F - 104 °F)

Warning: Inserting and connecting batteries incorrectly, short

circuiting or exposure to fluids pose an explosion hazard.

Warning: Do not mistreat, disassemble or attempt to repair the

battery.

Warning: Do not use the batteries if they are visibly damaged,

malfunctioning or appear to be leaking.

Warning: Take extreme care to avoid direct contact with electrolyte,

hot or smoking parts. In case of the above, disconnect and remove
the battery when it is judged safe to do so.

Using the Internal Compressor

The Patient Simulator’s chest movements, airway modes and
fluid systems are driven by compressed air. The right leg contains
a compressor and tank with separate reservoirs for clear and
simulated blood fluids.

For extended periods or stationary use, it is recommended to
connect to an external source of compressed air. This reduces wear
on the internal compressor and extends battery life of the Patient
Simulator.

For instructions on connecting an external compressor and adjusting
compressor default settings see

Air/CO2 Panel.

To avoid overheating and reduce wear

– If using the Patient Simulator in high temperatures, always allow

the simulator to cool down between training sessions.

Turning the Internal Compressor

OFF Using LLEAP

To switch OFF the internal compressor (to conserve the simulator
batteries and reduce wear), do the following:

1 In LLEAP, select the <Tools> menu.

2 From <Simulator Setup>, click <Turn off internal compressor>.

Changing Default Compressor

Settings Using LLEAP

Change default compressor settings via the Profile Editor.

1 Open the Profile Editor from the <Tools> menu in LLEAP.

2 Select the <Manikin hardware> tab in the Profile Editor.

3 From the compressor setup option, choose <Internal default>,

<External default> or <Remember last setting>.