Bleeding modules, Setup, Introduction – Laerdal SimMan 3G Trauma User Manual

Page 25: Removing simman 3g trauma left arm

Advertising
background image

25

SETUP

Bleeding Modules

Introduction

The SimMan 3G Trauma Bleeding Module kit includes trauma
modules which can be fitted to SimMan 3G Trauma to simulate
bleeding patient cases. After the simulation is completed, leave the
trauma modules connected, and perform the cleaning instructions as
stated in Maintenance.

– Amputated Arm Left SimMan 3G Trauma

– Amputated Arm Right SimMan 3G Trauma

– Amputated Leg SimMan 3G Trauma

– Amputated Lower Calf Skin

Removing SimMan 3G Trauma Left Arm

1 Unzip the clothing zippers on the right side. Remove the shirt.

2 Unzip the zippers on the left side of the torso. Open the Torso

Skin to one side.

3 Open the Stomach Foam to one side.

4 Lift the hinged chest plate to access the arm bolt.

5 Unscrew the left Arm Screw with the allen key and disconnect

all Arm cables.

LA PULSES

BP

LA

CONV

P

B

L

BP

PULSES

L

L

LA

B

6 Remove the arm with cables from the arm socket.

Note: Do not unscrew the Arm Screw fully.