DC-190 Ultra Count Installation Manual

3.0

Scale Setup

This section provides information about attaching scales and serial devices to the DC-190. Information presented
describes both physical connections and values that must be specified when configuring the DC-190 (see
Section 4.0). If you know what scales and serial devices will be connected to the DC-190, you can use the SPEC
Code Worksheets Section 10.1 on page 50 to record these values for later configuration.

3.1

Scale Configurations

The DC-190 can be connected to up to three analog scale bases using the load cell ports (PORT 1 and PORT 2
shown in Figure 3-1). An Ohaus

®

Explorer force balance can also connected to either the FORCE BALANCE or

232/PRINTER serial port for a total of four scale channels.

Figure 3-1. DC-190 Scale Ports and Connectors

Attach SX-L single and SX-D dual scale platforms to PORT 1 (left connector); attach external scale base to
PORT 2 (right connector). For three-scale configurations using either an SX-L and two external bases or three
external bases, an optional Y-cable must be used to connect two of the bases to PORT 1. (A wiring diagram of the
Y-cable is shown in Section 10.5 on page 55.)

3.1.1

Configuring SPECs 16 and 17

SPECs 16 and 17 are used to assign the annunciators for each of the four scale channels (

Scale No.

annunciator

on the DC-190 console).

All scale channels must be assigned, regardless of the actual number of scales attached.

Table 3-1 shows the values and default bit configurations for SPECs 16 and 17.

When configuring SPECs 16 and 17, use the following rules:

If a force balance is attached, it must be configured as Scale 1. That is, SPEC 16, bits 3 and 2, (Scale 1)
must be set to

Force Balance

(11).

Scales attached to Port 1 are configured as

Internal Scale 1

and

Internal Scale 2

, regardless of whether the

actual scale base is internal (SX-L, SX-D) or external.

SPEC

Default Setting

Bits 3 and 2

Bits 1 and 0

16

0001

Scale 1

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

Scale 2

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

17

1011

Scale 3

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

Scale 4

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

Table 3-1. Specifications 16 and 17 Default Settings