4 rs-232 specifications (143 settings) – Digi Counting Version 4.0 DC-190 User Manual

Page 19

Advertising
background image

24

DC-190 Ultra Count Installation Manual

4.4

RS-232 Specifications (143 Settings)

To configure RS-232 specifications, press and hold the

REZERO

key and enter

143

using the numeric keypad.

SPC40

appears in

WEIGHT

display and configuration of four bits (for SPEC 40) appears in

QUANTITY

display.

Follow the procedure described in Section 4.1 on page 17 to update the SPEC settings for the 143 access code.

SPECs 40–42 and SPEC 43, bits 0–2, represent the selectable RS-232 output fields.

SPECs 46–59 are not used. All bits for these SPECs must be set to 0.

SPEC

Bit 3

Bit 2

Bit 1

Bit 0

40

Lot Number

0: Yes
1: No

Memo (Part Name)

0: Yes
1: No

Alternative Part Number

0: Yes
1: No

ID Code

0: Yes
1: No

41

Net Weight

0: Yes
1: No

Gross Weight

0: Yes
1: No

Setpoint

0: Yes
1: No

Inventory

0: Yes
1: No

42

Date and Time

0: Yes
1: No

Quantity & Total Quantity

0: Yes
1: No

Unit Weight

0: Yes
1: No

Tare Weight

0: Yes
1: No

43

Autoprint within Setpoints

0: No
1: Yes

Non-stable Ouput

0: No
1: Yes (allow serial output
with scale in motion)

Status Output

0: Yes
1: No

Scale Number

0: Yes
1: No

44

Clear Operator Name
Each Use (BCP-300)

0: No (operator name held

in register for multiple
inputs)

1: Yes

Company Name
(BCP-300)

0: Default only
1: Input with bar board

Manual Print with Setpoints

0: Yes
1: No

Autoprint Overrange
Setpoint

0: No
1: Yes

45

Not used

Not used

Print with Zero Quantity
0: No
1: Yes

Batch Print Once

0: No
1: Yes

46–59

Not used: set to 0

Not used: set to 0

Not used: set to 0

Not used: set to 0

Table 4-3. DC-190 143 Settings