7 clearing locked-up scale (888888 error), Caution – Digi Counting Version 4.0 DC-190 User Manual

Page 31

Advertising
background image

Appendix

57

10.7 Clearing Locked-Up Scale (888888 Error)

Several configuration errors can result in the scale locking up with a display that shows all

888888

s. These errors

include the following:

If no force balance is connected to the DC-190, SPEC 8, bit 3, must be set to 0 (0xxx).

If a force balance is attached, SPEC 8, bit 3 must be set to 1 (1xxx) and SPEC 9, bit 2 (force balance
type), must be set to 1 (x1xx, Ohaus Explorer).

If no printer is attached to the DC-190, SPEC 10, bit 3, must be set to 0 (0xxx).

SPEC 16 and SPEC 17 must assign a different location for each scale, whether it is present or not. These
bits must be configured to show scale assignments for

Internal Scale 1

,

Internal Scale 2

,

External Scale

,

and

Force Balance

. See example below.

The following configuration for SPECs 16 and 17 contains duplicate entries for both Internal Scale 1
and
External Scale.

Do NOT set SPECs 16 and 17 to a configuration such as that shown below: a

lockup condition will result if this configuration is selected.

Should the DC-190 become locked up in all

888888

s as a result of invalid specification settings, an authorized

service technician can perform the following procedure:

1. Unplug scale from AC outlet.

2. Press and hold the

3

and

9

keys while plugging AC cord into outlet.

3. The display shows

S-on

. Release the

3

and

9

keys.

4. Press and hold

REZERO

key while keying access code

141

(display will show

SPC00

). Use the

+

key to

advance to the incorrect SPEC and edit its value.

5. When done, press the

*

key to store the new values, then press the

MODE

key to return to weighing

mode.

SPEC

Setting

Bits 2 and 3

Bits 0 and 1

16

0010

Scale 1
00: Internal Scale 1

Scale 2
10: External scale

17

1000

Scale 3
10: External scale

Scale 4
00: Internal Scale 1

Caution

Advertising