22

DC-190 Ultra Count Installation Manual

27

0100

Comma Display

0: No
1: Yes

Keypad Entry of Tare

0: No
1: Yes

Tare Range

00: 100% FS

01: 50% FS

10: 5% FS

11: Not used

28

0000

Auto Tare Clear when
Rezeroed

0:

No

1:

Yes (REZERO
clears tare value)

Automatic Unit Weight Clear Condition

00: Over net 5d and gross 21d, weight stable
01: Greater than or equal to net 1d, weight stable
10: Greater than or equal to net 1d, quantity > 0,

weight stable

Automatic Unit Weight
Clear

0: No
1: Yes

29

0000

Digital Tare Rounding

0: Tare exactly
1: Round to nearest

increment

Tare Value Exchange
with TARE key

0: Yes (allow tare

addition/subtraction)

1: No

Tare Addition

0: Yes (new tare weight

added to existing tare)

1: No

Tare Subtraction

0: Yes (new tare weight

subtracted from
existing tare)

1: No

30

none

Load Cell Sensitivity Selection (mV/V)—PORT 1, Scale 1
Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

Spec

Min

Max

0000:

3.46

4.00

0001:

3.00

3.46

0010:

2.59

3.00

0011:

2.25

2.59

Spec

Min

Max

0100:

1.95

2.25

0101:

1.69

1.95

0110:

1.46

1.69

0111:

1.27

1.46

Spec

Min

Max

1000:

1.09

1.27

1001:

0.95

1.09

1010:

0.82

0.95

1011:

0.71

0.82

Spec

Min

Max

1100:

0.61

0.71

1101:

0.53

0.91

1110:

0.46

0.53

1111:

0.40

0.46

31

none

Load Cell Sensitivity Selection (mV/V)—PORT 1, Scale 2
Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

Spec

Min

Max

0000:

3.46

4.00

0001:

3.00

3.46

0010:

2.59

3.00

0011:

2.25

2.59

Spec

Min

Max

0100: 1.95

2.25

0101: 1.69

1.95

0110: 1.46

1.69

0111: 1.27

1.46

Spec

Min

Max

1000:

1.09

1.27

1001:

0.95

1.09

1010:

0.82

0.95

1011:

0.71

0.82

Spec

Min

Max

1100:

0.61

0.71

1101:

0.53

0.91

1110:

0.46

0.53

1111:

0.40

0.46

32

1010

Calibration Mode
Protected by Span
Switch

0: Yes (span switch

must be pressed
before calibration)

1: No

Low Battery
Annunciator Enabled

0: Yes
1: No

Auto Exit from Part
Accumulation and
Reduction Mode

0: No (must press

CLEAR to perform
another accumulation)

1: Yes (exits to counting

mode after three
seconds)

Scale Connected to
PORT 2

0: No
1: Yes

CAUTION: If you are not
connecting a scale to this
connector, select 0.
Selecting 1 can damage
the console’s electronic
components.

33

0___

Over Weight Mask at:

0: +1d
1: +9d

Weight Decimal Point Position (PORT 2, Scale 2 or 3)
MUST BE SET PRIOR TO CALIBRATION!

000: 00000

011: 00.000

001: 0000.0

100: 0.0000

010: 000.00

101: 0.00000

34

0000

Not used

A/D Output (PORT 1,
Scale 1)

0: For std/normal load

cell (≤ 18 mV)

1: For abnormal load

cell with too large
offset (> 18mV)

A/D Filtering (PORT 1, Scale 1)

00: Normal
01: Protect from small vibration/fast change in

display

10: Protect from medium vibration
11: Protect from large vibration, slow change in

display

SPEC

Default

Bit 3

Bit 2

Bit 1

Bit 0

Table 4-2. DC-190 142 Settings (Continued)