Digi Counting Version 4.0 DC-190 User Manual

Page 15

Advertising
background image

20

DC-190 Ultra Count Installation Manual

15

0011

RS-232C Stop Bits
(Barcode pen)

0:

1 Bit

1:

2 Bits

RS-232C with Header
(Barcode pen)

0:

Yes (first character
recognized as
header)

1:

No (every bar code
treated as ID code)

RS-232C Parity Bit (Barcode pen)

00:

No

01: Odd

10:

Not used

11: Even

NOTE: All scale channels must be assigned in SPECs 16 and 17, regardless of the actual number of scales attached. See
Section 3.1.1 on page 12 for detailed information about configuring these SPECs.

16

0001

Scale 1

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

Scale 2

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

17

1011

Scale 3

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

Scale 4

00:

Internal Scale 1

01:

Internal Scale 2

10:

External Scale

11:

Force Balance

18

0000

Setpoint TTL Output

0:

Active low (0V)

1:

Active high (+5V)

Number of Setpoints. Values for 3–6 setpoints are valid only if SPEC 7, bits 0
and 1, are set to 10 or 11 (Quantity or Weight).

000: 2 setpoints
001: 3 setpoints
010: 4 setpoints
011: 5 setpoints
100: 6 setpoints

19

1000

Display “not F”
Message for Items not
Stored in Memory

0:

Yes

1:

No

Link to IMS (US version)

0: No
1: Yes

Type of Force Balance
(Japan version only)

0: SHG-300
1: HR-60

Print when Pressing

+

or

Key in Add Mode

0: Yes
1: No (print only with

*

)

SPEC

Default

Bit 3

Bit 2

Bit 1

Bit 0

Table 4-1. DC-190 141 Settings (Continued)

Advertising