Digi Counting Version 4.0 DC-190 User Manual

Page 18

Advertising
background image

Configuration Settings

23

35

0000

Not used

A/D Output (PORT 1,
Scale 2)

0: For std/normal load

cell (≤ 18 mV)

1: For abnormal load

cell with too large
offset (> 18mV)

A/D Filtering (PORT 1, Scale 2)

00: Normal
01: Protect from small vibration/fast change in

display

10: Protect from medium vibration
11: Protect from large vibration, slow change in

display

36

__00

Minimum Display (PORT 2, Scale 2 or 3)
MUST BE SET PRIOR TO CALIBRATION!

00: 2

01: 1

10: 5

11: 10

A/D Filtering (PORT 2, Scale 2 or 3)

00: Normal
01: Protect from small vibration/fast change in

display

10: Protect from medium vibration
11: Protect from large vibration, slow change in

display

37

1001

Load Cell Sensitivity Selection (mV/V)—PORT 2, Scale 2 or 3
Load cell sensitivity is automatically set at calibration. Manual adjustments can be made to reduce noise.

Spec

Min

Max

0000:

3.46

4.00

0001:

3.00

3.46

0010:

2.59

3.00

0011:

2.25

2.59

Spec

Min

Max

0100:

1.95

2.25

0101:

1.69

1.95

0110:

1.46

1.69

0111:

1.27

1.46

Spec

Min

Max

1000:

1.09

1.27

1001:

0.95

1.09

1010:

0.82

0.95

1011:

0.71

0.82

Spec

Min

Max

1100:

0.61

0.71

1101:

0.53

0.91

1110:

0.46

0.53

1111:

0.40

0.46

38

0010

A/D Output (Scale 3)

0: For std/normal load

cell (≤ 18 mV)

1: For abnormal load

cell with too large
offset (> 18mV)

Digital Tare with Weight
on Scale

0: Yes
1: No

Internal Count

0:

500,000

1:

1,000,000

Stability Check When
Changing Scale

0: Yes
1: No

39

0010

Set SPEC 39 to 0010

SPEC

Default

Bit 3

Bit 2

Bit 1

Bit 0

Table 4-2. DC-190 142 Settings (Continued)