• Do not use detergents or solvents.

• On no account must you immerse the computer in water,

otherwise liquid can enter it and cause demage.

• When storing the device, make sure that no heavy objects

are placed on top of it. Remove the batteries. The cuff tube

should not have any sharp kinks.

8. Rectifying faults

Error messages can occur when

• you move or talk during the measurement (

,

),

• the cuff tube is not properly inserted (

),

• when changing the batteries, the start/stop button was

pressed,

• inflation takes longer than 20 seconds (

),

• the inflation pressure is higher than 300 mmHg (

).

In the above cases, you must repeat the measurement. Make

sure that the cuff tube is properly inserted and that you do

not move or talk. Re-insert the batteries if necessary, or else

replace them.

9. Specifications

Model no.

BM 60

Measurement

method

Oscillometric, non-invasive blood pres-

sure measurement on the upper arm

Measurement range

Cuff pressure 0 – 300 mmHg,

systolic 50 – 250 mmHg,

diastolic 40 –180 mmHg,

Pulse 40 –160 beats/minute

Display accuracy

Systolic ± 3 mmHg, diastolic ± 3 mmHg,

pulse ± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation

according to clinical testing:

systolic 8 mmHg /diastolic 8 mmHg

Memory

2 x 60 memory spaces

Dimensions

L 101 mm x W 130 mm x H 54 mm

Weight

Approx. 387 g (without batteries)

Cuff size

22 to 32 cm

Permissible operating

conditions

+10 °C to + 40 °C, ≤ 85 % relative

air humidity (non-condensing),

700 –1060 hPa ambient pressure

Permissible storage

and transport con-

ditions

- 5 °C to + 50 °C, ≤ 85 % relative air

humidity

Power supply

4 x 1,5 V

AA batteries

Battery life

For approx. 300 measurements,

depending on the blood pressure level

and/or pump pressure

Accessories

Instruction for use, 4 x 1.5 V AA batter-

ies, storage pouch

Classification

Internal supply, IPX0, no AP or APG,

continuous operation, application part

type BF (cuff), CLASS II (when using a

mains part)

Maximum surface

temperature

Device: 44,1 °C

Mains part: 47,2 °C