Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity
BM 60 is intended for use in the electromagnetic environment specified below. The customer or the user of BM 60 should assure that it is used

in such an environment.
Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If

floors are covered with synthetic material, the rela-

tive humidity should be at least 30 %.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

±2 kV for power supply lines

±1 kV for input/output lines

Mains power quality should be that of a typical

commercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

± 1 kV differential mode

± 2 kV common mode

Mains power quality should be that of a typical

commercial or hospital environment.

Voltage dips, short interrup-

tions and voltage variations

on power supply input lines

IEC 61000-4-11

< 5 % U

T

(> 95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

< 5 % U

T

(> 95 % dip in U

T

)

for 5 sec

< 5 % U

T

(> 95 % dip in U

T

)

for 0.5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

< 5 % U

T

(> 95 % dip in U

T

)

for 5 sec

Mains power quality should be that of a typical

commercial or hospital environment. If the user of

BM 60 requires continued operation during power

mains interruptions, it is recommended that BM 60

is be powered from an uninterruptible power supply

or a battery.

Power frequency (50/60 Hz)

magnetic field IEC 61000-4-8 3 A/m

3 A/m

Power frequency magnetic fields should be at levels

characteristic of a typical location in a typical com-

mercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.