Beurer GL40 mg/dL

1

Twist to remove the cap from the lancing device.

2

Place a sterile needle lancet in the lancing device and push

the lancet firmly into place.

3

Remove the protective disc from the lancet by twisting while

holding the shaft of the lancet firmly. Retain the protective

disc so that you can safely dispose of the used needle lancet

after taking the blood sample.

4

You will need to use a different cap depending on the site

from which you want to take the sample:

Fingertip: Cap

Other body parts: AST cap (transparent)

Place the selected cap on the lancing device and twist it

into place.

5

Setting the puncture depth:

You can preset six different puncture depths on the lancing

device by turning the cap. A small bar means a shallow

puncture depth, and a large bar means a deep puncture

depth.

The difference in puncture depth between the smallest and

largest click position is approx. 2 millimetres

• Small bar (1 to 2): Delicate or thin skin

• Medium bar (3 to 4): Normal skin

• Large bar (5 to 6): Thick or rough skin

Turn the cap to the required click position until the arrow in-

dicates the desired puncture depth.