Beurer GL 40 (mg/dL) User Manual

Page 27

Advertising
background image

Beurer GL40 mg/dL

27

Test strip inserted incorrectly or not completely. Insert the test strip, contacts first, into the slot

on the unit. Ensure that the front side of the test

strip is facing you (see "Test strips", page 12).

Unit is defective

Contact customer service.

Problem: After inserting the test strip into the unit and applying the blood, the test does not start.

Cause

Solution

Insufficient quantity of blood

Repeat the test using a new test strip and larger

blood drop.

Test strip is defective

Repeat the test using a new test strip.

Blood was applied while the device was switched

off.

Repeat the test using a new test strip and only

apply blood after and flash.

The unit's basic settings were changed and the

change was not completed (see "Basic set-

tings", page 14).

Press the "On/Off" button until “ ” is dis-

played.

Repeat test.

Unit is defective

Contact customer service.

9 t

echnicAl

specificAtions

Dimensions (W x H x D)

47 x 85 x 14 mm

Weight

43 g incl. batteries

Power supply

2 x 3 V CR 2032 button cell batteries

Battery life

More than 1000 measurements

Measured value memory

480 measured values with date/time

Data retained upon changing batteries

Average values

for 7, 14, 30, 90 days

Automatic switch-off

Two minutes after last actuation

Storage/

transport temperature

Temperature: +2 °C – +30 °C

Relative humidity: < 90 %

Operating ranges

Temperature: +10 °C – +30 °C

Relative humidity: < 90 % non-condensing

Measuring range, glucose

Glucose: 20 - 630 mg/dL

Blood sample

capillary whole blood

Amount of blood

0.6 microlitres

Blood glucose measuring duration

Approx. 5 seconds

Advertising