Note

The measuring device is made from precision components. The accuracy of the measured values

and the service life of the unit depend on careful handling:

• Protect the unit against impact and do not drop it.

• Protect the unit against damaging influences such as moisture, dirt, dust, blood, control solution or

water, extremes of temperature and direct sunlight.

• Do not store close to strong electromagnetic fields, wireless systems or mobile telephones.

8 t

roubleshootinG

Messages on the display for batteries and blood glucose measurement

No.

Cause

Solution

LP

Flat batteries

Replace all batteries.

Ht

Temperature of the measurement room,

measuring device or test strip above the

allowable limit.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room

temperature (+20 °C to +25 °C).

Lt

Temperature of the measurement room,

measuring device or test strip below the

allowable limit.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room

temperature (+20 °C to +25 °C).

Err

Used or contaminated test strip was in-

serted.

• Insert unused test strip that has not ex-

pired.

• Repeat the blood glucose measurement.

Err001 System error.

Remove batteries, reinsert batteries. Contact

our customer service if the problem persists.

Err005 Memory error.

Remove batteries, reinsert batteries. Contact

our customer service if the problem persists.

Problem: Unit does not turn on.

Cause

Solution

Flat batteries

Replace batteries.

Incorrectly inserted or missing batteries.

Verify that the batteries are inserted correctly

(see "Inserting and replacing the batteries",

page 13).