Beurer GL 40 (mg/dL) User Manual

Page 28

Advertising
background image

28

Beurer GL40 mg/dL

Calibration

Plasma

Test procedure

Amperometric biosensor

Using the device

Suitable for private, individual use

System function test

Each time the unit is switched on

EMC
This unit complies with European standard EN 61326 and is subject to specific safety precautions

in terms of electromagnetic compatibility. Note that portable and mobile RF communication equip-

ment can affect this unit. Further details can be requested by contacting the customer service ad-

dress provided below.

Note regarding function of test strips
The test strips enable a quantitative measurement of the glucose content of capillary whole blood.

When the slit for collecting blood comes into contact with a drop of blood, it fills automatically through

simple capillary action. The blood is sucked into the absorbing slit of the test strip and the measuring

device measures the glucose level of the blood.

The test is based on the measurement of an electric current caused by a chemical reaction between

the glucose and the enzyme glucose oxidase (aspergillus niger) on the test strip. During the reaction,

a mediator transports electrons through the electrode surface, thereby generating a current. The

measuring device analyses this current. The flow of current is proportional to the glucose content of

the blood test. The results are shown in the display of the measuring device. Only a small quantity of

blood is required (0.6 microlitres) and the measurement takes approx. five seconds. The strips can

detect blood glucose levels in the range 20 – 630 mg/dL.

Chemical components of the test strip sensor
• Glucose oxidase ≥ 0.6 IU

• Electron shuttle ≥ 0.03 mg

• Non-reactive components ≥ 0.06 mg

Note regarding function of the control solution
The control solution contains a fixed concentration of glucose that reacts with the test strip. A test

with the control solution is performed in the same way as a blood test, but the control solution is

used instead of blood. The measurement result from the control solution must lie within the results

range. This results range is printed on every test strip container.

Chemical composition
The control solution is a red solution with the following D-glucose level (in percentage shares):
Ingredients

Control solution LEVEL 1

Control solution LEVEL 2

D-glucose

0.10 %

0.35 %

Non-reactive components

99.90 %

99.65 %

Standards
The Beurer GL40 measuring system meets the following European directives and standards: