Beurer GL 44 (mg/dL) User Manual

Page 32

Advertising
background image

32

Beurer GL44 mg/dL

8 w

hat

if

there

are

problems

?

Display messages on batteries and blood glucose measurement

No.

Cause

Solution

LP

Batteries empty.

Replace all batteries.

Ht

Temperature of the measuring environ-

ment, blood glucose monitor or test strip

above the permitted range.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room

temperature (+20 °C to +26 °C).

Lt

Temperature of the measuring environ-

ment, blood glucose monitor or test strip

below the permitted range.

Repeat the test using a new test strip as soon

as the measuring environment, blood glucose

monitor and test strips have reached room

temperature (+20 °C to +26 °C).

Err

Used or contaminated test strip inserted. • Insert unused test strip that has not ex-

pired.

• Repeat the blood glucose measurement.

Err001 System error.

Remove batteries, reinsert batteries. Contact

our customer service if the problem persists.

Err002 Insufficient amount of blood on the test

strip.

Repeat the measurement using a new test

strip.

Err003 The hematocrit value is outside the range

20 to 60%

Repeat the measurement using a new test

strip. Contact Customer Services if the prob-

lem persists.

Err005 System error.

Remove batteries, reinsert batteries. Contact

our customer service if the problem persists.

Unknown error messages.

Remove batteries, reinsert batteries. Contact

our customer service if the problem persists.

Problem: device does not switch on

Cause

Solution

Batteries empty.

Replace batteries.

Incorrectly inserted or missing batteries.

Check whether the batteries have been inserted

correctly (see “Inserting and replacing the bat-

teries” page 14).