Standards
The Beurer GL44 monitor corresponds to European guidelines IVD (98/79/EC) and MDD (93/42/EC).

10 c

omparison

of

measured

values

with

laboratory

value

Precision
Three lots of the GL44 blood glucose test strips have been tested to assess the precision of the GL44

blood glucose measuring system. This includes a repeat assessment using venous blood and a labora-

tory precision assessment using the control material. The blood glucose content of the venous blood

samples ranges from 42.7 to 418.0 mg/dL and control material from three concentrations is used.

Results of the repeat precision measurements

Sample

Venous blood

(mg/dL)

Grand mean

value (mg/dL)

Pooled standard de-

viation

Pooled coefficient

of variation (%)

1

42,7

36,0

2,0

5,6

2

62,0

59,2

3,5

5,9

3

120,5

127,1

4,1

3,2

4

201,0

213,8

6,7

3,1

5

316,5

329,9

10,1

3,1

6

418,0

433,5

14,5

3,3

Results of the intermediate precision measurement

Sample

Control material

(mg/dL)

Grand mean

value (mg/dL)

Pooled standard

deviation

Pooled coefficient

of variation (%)

1

70,0

71,3

1,0

1,4

2

135,6

136,3

1,4

1,1

3

351,5

350,8

2,8

0,8

System accuracy
The GL44 blood glucose monitor in comparison with the YSI.

Three lots of GL44 blood glucose test strips have been tested to assess the system accuracy of the

GL44 blood glucose measuring system and to compare it with the reference method in which capil-

lary whole blood concentrations of 34.4 to 442.8 mg/dL have been used.