Beurer GL 44 (mg/dL) User Manual

Page 33

Advertising
background image

Beurer GL44 mg/dL

33

Cause

Solution

Test strip inserted incorrectly or not completely. Firmly insert the test strip into the slot on the

device with the contacts first. Make sure that

the front of the test strip is facing you (see “Test

strip” on page 12).

Device faulty.

Contact customer service.

Problem: the test does not start after inserting the test strip into the device and applying blood

Cause

Solution

Insufficient amount of blood or test strip not

filled correctly.

Repeat test using a new test strip and a larger

drop of blood. Please note the correct filling of

the test strip (see page 20).

Faulty test strip.

Repeat the test using a new test strip.

Blood was applied while the device was switched

off.

Repeat the test using a new test strip and only

apply blood after and flash.

The basic settings of the device were changed

and the changes were not completed (see “Mak-

ing and changing basic settings” on page 14).

Remove the test strip and press the “On/Off”

button until “

” is displayed.

Repeat test.

Device faulty.

Contact customer service.

9 t

echnical

specifications

Dimensions (W x H x D)

52 x 95 x 16 mm

Weight

44 g incl. batteries

Power supply

2 x 3 V CR 2032 button cell batteries

Battery life

500 measurements with backlighting

1000 measurements without backlighting

Measured value memory

480 measured values with date/time

Data retained when batteries are changed

Average values

for 7, 14, 30, 90 days

Automatic switch-off

Two minutes after last actuation

Storage/

transport temperature

Temperature: +2 °C – +30 °C

Relative humidity: < 90 %

Operating ranges

Temperature: +10 °C – +40 °C

Relative humidity: < 90 % non-condensing

Measuring range, glucose

Glucose: 20 – 630 mg/dL