Fluke Biomedical 945A User Manual

Page 24

Advertising
background image

Set-up

To place the UDR in operation, the following steps should be performed:

NOTE

Ensure you have read and fully understand section 4, prior to continuing.

1. Verify jumpers and DIP switches on the preamplifier and ratemeter are set for the operational

features desired. Refer to Section 4 Tables 4-3 through 4-6 for additional information.


2. Apply power to the unit.

3. Enter detector dependent set points such as analog output, High/Low scale Value, and calibration

constant, using the procedure outlined in Section 4, and Tables 4-2 & 4-7.


4. Enter channel dependent setpoints such as Alarm Limits are to be entered using the procedure

described in Section 4, and Tables 4-2 & 4-7.


5. Electronic adjustments affecting calibration are factory set and should not be readjusted for initial

operation unless the UDR has been in storage for more than one year. If the Unit has been in
storage for more than one year, the electronic calibration described in Chapter 6 should be
performed.