Preamplifier Interface Board 977-210-10 Switch Settings

Switch

Normal Op. Switch Function

SW3-1

ON

0/OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)

SW3-2

OFF

0/OFF Selects alarm inhibit, SW3-3 must be on for this state to be true.
1/ON Selects alarm enable

SW3-3

OFF

0/OFF Selects Rate Mode (normal) • 1/ON Selects Maintenance Mode

SW3-4

OFF

0/OFF forces 977 analog output to 0 V (low scale)
(SW3-3 must be on for state to be true)
1/ON forces 977 analog output to 10 V
(SW3-3 must be on for state to be true)

NOTE

The above switches are located in a 4 position dip switch on the
977 interface board (977-210-10).


Jumper Positions for Preamplifier Interface Board 977-210-10

Jumper Normal

Setting

Function Alternate

Setting

Function

JMP1

A to B

Enables Victoreen Loop (normal)

B to C

Enables RS232C

JMP2

N/A

Deleted, Not Used

N/A

Deleted, Not Used

JMP3

A to B

Enables Alarm Output -
Non-Failsafe

B to C

Enables Alarm Output
Failsafe

JMP4

A to B

Enables 0 - 10 Vdc Analog Output
(normal)

B to C

Disables Analog Output (stays at
10 Vdc)

JMP5

B to C

Enables HV Output (normal)

A to B

Disables HV Output

Jumper Positions for Preamplifier Electrometer Board 977-200-15

Jumper Normal

Setting

Function Cal.

Position

Function

JMP1

A to B

Normal Operation

B to C

PGA Offset Adjust *

* Programmable Gain Amplifier

Model 946A-200 Option Jumpers

Jumper Function

Norm

Position

Operation

JP3-1,

Not Defined

in

Not Applicable

JP3-2

Not Defined

in

Not Applicable

JP3-3

Alarm Reset Select

in

Manual reset with jumper installed
Auto reset with jumper removed

JP3-4

Fail Alarm Enable

in

Fail alarm enabled with jumper in
Fail alarm disabled with jumper out

JP3-5

Check Source Alarm Mute in

Alarms disabled during C/S with
jumper in
Alarms enabled during C/S with
jumper out

8-5