Udr function switch / function -17, Rror, Odes – Fluke Biomedical 945A User Manual

Page 41

Advertising
background image

UDR Function Switch / Function

Table 4-7 shows the UDR Function Switch positions and their function.

Table 4-7. UDR Function Switch Positions/Function

Switch
Position

Function

0

Selects HIGH Alarm Setpoint

1

Selects WARN Alarm Setpoint

2

Selects Overrange Limit

3

Selects Underrange Limit

4

Selects Analog Full Scale

5

Selects Calibration Factor

7

Selects Analog Low Scale

8

Selects Auto Zero Display Mode

9

Selects Loop Test Mode



Error Codes

The error code for the 946A-200 UDR are listed in Table 4-8.

Table 4-8. Model 946A-200 Error Codes

Error Codes

Function

Reference Section

E0001

Negative display data

Setpoint Error Codes

E0002

Invalid setpoint value (s)

Setpoint Error Codes

E0007

Loop time out

Communication Loop Failure

E0008

Invalid analog scale values

Setpoint Error Codes

E0009

Loop test failure

Loop Test Mode

E0011*

High voltage failure

Fail Alarms

E0012*

Auto zero correction failure

Fail Alarms


* These error codes always indicate a detector failure.

4-17