The UDR contains a series of five hardware jumpers and up to seven user specified set points which
affect operation of the unit.

The jumpers are designated as JP3-1 through JP3-5. Refer to the main circuit board assembly drawing
located in Appendix B of this manual for JP3 locations. The functions affected by JP3 are described in
the following paragraphs.

UDR Switch & Table Settings

Jumper and switch settings for setpoint selection for the UDR are summarized in Tables 4-6 and 4-7,
while error codes are listed in Table 4-8.

Jumper Selectable Options

NOTE

The unit must be turned off when changing jumpers.


Refer to Table 4-6 for a list of option jumpers and their functions.

With jumper JP3-3 (Alarm Automatic Acknowledge) installed, the operator is required to acknowledge the
HIGH and WARN dose rate alarms. With the jumper removed, the HIGH and WARN alarm will be
automatically acknowledged and will be cleared when the dose rate falls below the setpoint.

With jumper LP3-4 (Enable Fail Alarm) installed, the 5-minute low signal Fail Alarm is enabled. With the
jumper removed, the 5-minute low signal Fail Alarm is disabled. Note that other conditions which cause a
Fail Alarm are not affected by this jumper setting.

Table 4-6. Model 946A-200 Option Jumpers

Jumper Function

Norm

Position

Operation

JP3-1,

Not Defined

in

Not Applicable

JP3-2

Not Defined

in

Not Applicable

JP3-3

Alarm Reset Select

in

Manual reset with jumper installed
Auto reset with jumper removed

JP3-4

Fail Alarm Enable

in

Fail alarm enabled with jumper in
Fail alarm disabled with jumper out

JP3-5

Check Source Alarm Mute in

Alarms disabled during C/S with
jumper in
Alarms enabled during C/S with
jumper out