Fluke Biomedical 945A User Manual

Page 79

Advertising
background image

Connector P2 - Detector Connector

Pin Signal

Internal

Connection

1

+15 VDC

Not used in most applications

2

-15 VDC

Not used in most applications

3

Supply Ground

Power Supply Ground

4

Transmit +

Communication Board

5

Transmit -

Communication Board

6

Receive +

Communication Board

7

Receive -

Communication Board

8

Loop Voltage +
Fused + 15 VDC

Communication Board

9

Loop Voltage -
Fused -15 VDC

Communication Board

10

0 - 10 Vdc*

Main Circuit Board J5-10

11

Ground

Main Circuit Board J5-9

12

Check Source Line

Not used in most applications

13

Neutral

Not used in most applications

14

Alarm Switch Line

Not used in most applications

* 0 - 10 Vdc selected for customer use on P2 (10,11) or P6 (5, 6), not both.

Connector P3 - Power Input

Pin Signal

Internal

Connection

1

120 Vac, Line

Line fuse (F2)

2

120 Vac, Neutral

Power Supply, 120 Vac (n)

3 Safety

Ground

Chassis

8-3