Fluke Biomedical 945A User Manual
Page 79

Connector P2 - Detector Connector
Pin Signal
Internal
Connection
1
+15 VDC
Not used in most applications
2
-15 VDC
Not used in most applications
3
Supply Ground
Power Supply Ground
4
Transmit +
Communication Board
5
Transmit -
Communication Board
6
Receive +
Communication Board
7
Receive -
Communication Board
8
Loop Voltage +
Fused + 15 VDC
Communication Board
9
Loop Voltage -
Fused -15 VDC
Communication Board
10
0 - 10 Vdc*
Main Circuit Board J5-10
11
Ground
Main Circuit Board J5-9
12
Check Source Line
Not used in most applications
13
Neutral
Not used in most applications
14
Alarm Switch Line
Not used in most applications
* 0 - 10 Vdc selected for customer use on P2 (10,11) or P6 (5, 6), not both.
Connector P3 - Power Input
Pin Signal
Internal
Connection
1
120 Vac, Line
Line fuse (F2)
2
120 Vac, Neutral
Power Supply, 120 Vac (n)
3 Safety
Ground
Chassis
8-3