Preamplifier/UDR Diagnostic LED Indicators

977-210 Detector/Preamplifier

The LEDs listed below are located on the 977-210-10 Communication/Interface board. This is the larger
of the two circuit boards in the detector assembly.

Designator

Function / Status Indicated

LED 1

+15 VDC is being supplied to the detector from the
UDR when ON

LED 2

ON (bright): both + 15 and -15 volt loops present
ON (dim): Only one loop supply present
OFF: no loop voltage from either supply

LED 4

HV is being generated when LED is ON. No HV
supply when LED is OFF

LED 5

+12 VDC is being supplied to the HV driver when
LED is ON.
+12 VDC is not present to the HV driver when LED
is OFF


NOTE: There is no LED 3 installed on the circuit board

946A-200 UDR

The sole diagnostic LED in the UDR is CR10, located on the 942-200-80 Communications Interface
Board. When CR10 is ON (bright), both the +15 and -15 loop voltages are present at receive. A dim
CR10 indicates only one supply is present. CR10 is OFF indicates that no loop supplies are present at
the receive terminal.

4-18