Fluke Biomedical 945A User Manual

Page 56

Advertising
background image

Relay Control Register (Write Only)

The Relay Control Register (U44) is an 8 bit register with clear, and responds to address 400C. Upon
initial power-up, the system reset signal sets all outputs low. Data written into U44 remains at the outputs
until a reset occurs or new data is written. Table 4-15 lists relay control write functions.

Table 4-15. Relay Control Register Write Functions

Data

Bit

Function

D0

Fail Relay and Indicator

D1

Check Source Relay

D2

Warn

Relay

D3

Alarm

Relay

D4

Rate-of-rise Relay (not used)

D5

Remote

Indicator


With the exception of the fail bit (D0), all outputs are applied to U48, which is an inverting open collector
driver. The outputs of U48, including fail, drive (via the J2 connector) mechanical relays located on the
relay board. The fail bit is used as an input to U47-3 which, when set high-low-high once per second,
causes output U47-6 to remain low. Should this high-low-high sequence fail to occur (under MPU
control), U47 will time out and set U47-6 high, causing the fail indicator on the panel and the fail relay to
de-activate.

Bargraph (Write Only)

The bargraph addresses are decoded by U5, which is a 1 of 8 decoder. Control line inputs to U5 are
R/W, Write 02, and Bargraph. When the bargraph address is selected (4020-4025), U5 further decodes
the addresses listed in Table 4-16.

Table 4-16. Bargraph Write Address


Address State

Hex Address

Function

Driver

A3 A2 A1

0 0 0

4020

Green

1-10

U55,

U56

0 0 1

4021

Green

11-20

U57,

U58

0 1 0

4022

Green

21-24

U59

0 1 1

4023

Red

1-10

U50,

U51

1 0 0

4024

Red

11-20

U52,

U53

1 0 1

4025

Red

21-24

U54

4-32