Fluke Biomedical 945A User Manual

Page 45

Advertising
background image

Power for the detector is normally provided by the electronic high voltage power supply ( jumper 5 in
position B to C). However, high voltage can be supplied to the detector by connecting battery terminals 1
(+) and 2 (-) of J7, and placing jumper 5 in the A to B position.

The high voltage board has a 100 to 1 voltage divider circuit which permits monitoring high voltage
through operational amplifier U15 on TP5 with an ordinary voltmeter (10 Meg impedance), without loading
down the high voltage circuit or requiring the use of a high impedance probe. The high voltage
adjustment is provided by R118. Refer to section 5 for adjustment instructions and precautions.

The following LEDs are provided to indicate operation of certain functions of the 977-210-10 board:

LED #

Indicates

LED1

The presence of +15 Vdc supplied to the
circuit board from the UDR

LED2

Communications - Receive

LED4

Regulated +12 Vdc is supplied to the
electronic high voltage circuit

LED5

High voltage is present at J6



Test points are provided on the high voltage board as follows:

Test Point

Function

TP3

High Voltage 1:1

TP4 DC

Ground

TP5

High voltage divided by 100

NOTE

TP1 and TP2 are not installed on the 977-210-10 circuit board.


A 0 - 10 Vdc analog output signal is obtained from an 8 bit D/A converter U7 and operational amplifier
U11. JMP4 controls the operation of the analog output. This output is normally used to drive a local
meter or display. The analog output is logarithmic.

Transistors Q13, Q101, Q103, and Q102 are used to provide a -15 Vdc output voltage to drive a local
alarm. The normal state of the output (energized or de-energized) is selectable via jumper JMP3.

4-21

Advertising