Data Entry (Read Only)

Octal buffer U36 functions as an interface to supply the status of the switches for data entry to the internal
data bus address (4002). The data entry read functions are shown in Table 4-20. The function switch
logic is shown in Table 4-21

Table 4-20. Data Entry Read Functions

Data

Bit

Function

D0

Enter Switch (Active Low)

D1

Value Select Switch (Active Low)

D2

Digit Select Switch (Active Low)

D3

(Not

Used)

D4

(16 position function switch, see Table 4-21)

D5

(16 position function switch, see Table 4-21)

D6

(Not

Used)

D7

(Not

Used)

Table 4-21. Function Switch Logic


Hex

Data

D7 D6 D5 D4

Position

Function

F

1 1 1 1

0

High

E

1 1 1 0

8

Auto

Zero

Test

D

1 1 0 1

1

Warn

C

1 1 0 0

9

Loop

Test

Mode

B

1 0 1 1

4

Analog

Full

Scale

Value

A

1 0 1 0

C

Undefined

9

1 0 0 1

5

Calibration

Factor

8

1 0 0 0

D

Undefined

7

0 1 1 1

2

Overrange

6

0 1 1 0

A

Undefined

5

0 1 0 1

3

Underrange

4

0 1 0 0

B

Undefined

3

0 0 1 1

6

Undefined

2

0 0 1 0

E

Undefined

1

0 0 0 1

7

Analog

Low

Scale

Value

0

0 0 0 0

F

Undefined


Sensitivity Select (Read Only)

Octal buffer U42 functions as an interface to provide the status of the sensitivity selection jumpers, mode

jumpers, anti-jam bit data, and serial data from E2 PROM to the internal data bus and MPU. U42
responds to address 4008. The sensitivity read functions are listed in table 4-22.