Section 6 - Calibration

Calibration

The 977 Ion Chamber Detector is factory calibrated on a 137Cs/60Co range. A ± 15%, 10 point NIST
traceable factory calibration is performed to obtain a calibration factor for the detector. The calibration
factor is stated on the factory calibration data sheet, and is adsjusted via set-point 5 on the 946A UDR. A
field test source, P/N 848-5, is available as an option to verify detector operation. The
detector/preamplifier pair should be returned to Fluke Biomedical RMS for recalibration whenever
performance degredation is observed.

Electronic Adjustments

Table 6-1 lists the electronic adjustments for the 946A UDR and 977-210 preamplifier. The procedures
for making these adjustments are contained in the factory Calibration Procedure listed below. Note that
test firmware is required to perform the electrical adjustments. Refer to Section 7 for a description of user
permitted adjustments.

Table 6-1. Electronic Adjustments

Adjustment

(UDR)

Purpose

Tolerance

VR1

Zero Adjust / 0 - 10 Volt Analog Output

0.000 Vdc ±5 mV

VR2

Gain Adjust / 0 - 10 Volt Analog Output

±.5% of full scale

VR4

20 mA Adjust / 4 - 20 mA Volt Analog Output #1

±.5% of full scale

VR5

20 mA Adjust / 4 - 20 mA Volt Analog Output #2

±.5% of full scale

VR6

4 mA Adjust / 4 - 20 mA Volt Analog Output #2

±.5% of full scale

VR7

4 mA Adjust / 4 - 20 mA Volt Analog Output #1

±.5% of full scale

VR13

Write Cycle Clock Adjust

±5 ns

Adjustment

(977)

Purpose

R11

Programmable Gain Amplifier Offset Adjustment

± 2 mV

R12

Auto Zero Set-Up Adjustment

± 10 mV

HV Supply R118

High Voltage Adjust

± 20 V

R53

Analog Output Adjustment

± 10 mV

Note

Refer to the drawings in Appendix B to locate adjustment and/or test points, if required.

WARNING

Sections of this circuit board contain voltages in excess of 500
volts. Use extreme care when making any adjustments.

Use extreme care when making high voltage measurements directly
at J6 or TP3.

The follow factory calibration and test procedures are included for reference.

Document

Description

Cal-ION2

Factory Calibration Procedure for the 977 Detector

TP977-201/210

Test Procedure for 977 Series

LT946/977

Factory Loop Test for 945 System

TP946A-100-3

Factory Test Procedure for 946A-200

MB95-1004

Field Cable Termination for Replacement 977-210 Detector

TP-977 Cable

Factory Test Procedure for 977-210-66 cables

6-1