Please use only the original
standard programs, which are
especially coordinated with the
corresponding materials for the
Ivoclar Vivadent all-ceramic
systems (IPS

e.max

, IPS Empress).

19

Freely programmable press programs
Furthermore, the EP 600 is equipped with
4 freely programmable press programs
(P8–P11). For these programs, values within
the following ranges can be entered:

Important information
For the layering technique, we
recommend a stop speed of
300 mm/min, for the staining
technique 150 mm/min.

Higher values (stop speed of e.g.
300 mm/min) results in the press cycle
being stopped sooner.

Lower values (stop speed of e.g.
100 mm/min) results in the press cycle
being stopped later. Consequently, the
press cycle is prolonged.

Once the program is started by pressing
START, the new program is displayed in the
usual manner.

Parameter

B Stand-by temperature

t Temperature increase

T Holding temperature

H Holding time

E Stop speed

MIN

50

1

50

0

0

MAX

900

140

1200

109

1000

Unit of measure

°C

°C/min.

°C

min.

µm/min