Product specification – Ivoclar Vivadent EP 600 User Manual

Page 30

Advertising
background image

30

9.2 Technical data

Power supply:

200–240 Volt, 50–60 Hz
110–120 Volt, 50–60 Hz

Installation category II

Max. power consumption:

12 A at 110–120 VAC

8,5 A at 200–240 VAC

Vacuum quality:

100% 40–120 mbar

Acceptable data for pumps of other manufacturers:

Max. power consumption: 2,1 A

Final vacuum: < 50 mbar

Use only tested pumps!

Electrical fuses:

200–240V: T 6,3 A (heating circuit) T 1 A (controls)

T3,15 A (vacuum pump)

110–120V: T 12 A (heating circuit) T 2 A (controls)

T5 A (vacuum pump)

Fuse dimensions:

200–240V = diameter 5 x 20 mm

110–120V = diameter 6,3 x 32 mm

Dimensions:

w x d x h: 445 x 520 x 650 mm

Usable size of the press chamber:

diameter 50 mm, height 85 mm

Max. press temperature:

1200 °C

Weights:

Furnace head: 7,6 kg

Furnace base: 17,0 kg

Tray: 0,4 kg

Press furnace, complete: 25,0 kg

Extra muffle: 0,3 kg

Safety information:

IEC 1010-1, EN 61010-2-020, Part 1

ULC and cUL standards

Radio protection, electromagnetic compatibility:

EMC tested

This chapter contains all the
relevant product specifications.

9.1 Delivery form

EP 600
– EP 600 incl. software
– Sagger tray
– Automatic Temperature

Checking Set 1 (complete)

– Investment ring cooling grid
– Power cord
– Vacuum hose
– Operating Instructions
– Firing plate (slit)

Accessories
– Refill package ATK1 "Melting

Samples"

– Refill package ATK 1 "

Ceramic Base Block"

– VP3 Vacuum Pump

Colours
Standard colour:
Traffic white (RAL 9016)

Special colours:
– Salmon (RAL 3014)
– Aquamarine (RAL 5014)
– Turquoise (RAL 6027)

9. Product Specification

Advertising