7. Maintenance, Cleaning, Diagnosis

This chapter describes the user maintenance and cleaning proce-
dures for the IvoBase Injector. Only those tasks are listed that may
be performed by dental professionals. All other tasks must be per-
formed by qualified service personnel at a certified Ivoclar Vivadent
Service Centre.

The time for these maintenance procedures depends on the fre-
quency of use and the working habits of the users. For that reason,
the recommended times are only approximations.

This injector has been developed for typical use in dental

laboratories. If the product is used for continuous opera-
tion, premature ageing of the expendable parts has to
be expected.

Expendable parts are, for instance:
– Heater
– Clasps

Expendable parts are not covered by the warranty.
Please also observe the shorter service and maintenance intervals.

7.1. Monitoring and maintenance

What:

When:

Check all plug-in connections for correct fit.

weekly

Check the water level in the waste water con-
tainer and empty it, if necessary (see 7.2 for
details).

daily

Check if the heater is dirty or damaged (see 7.3
for details).

weekly

Check if the temperature sensor in the poly-
merization chamber is dirty or damaged (see
7.4 for details).

weekly

Check if the safety door is dirty or damaged.

daily

Check the keypad for visible damage. If the key-
pad is damaged, it has to be replaced by a
certified Ivoclar Service Centre.

weekly

7.2 Emptying the waste water container

During the polymerization process, condensation develops within
the injector, which is discharged via a drain outlet. Check the water
level at regular intervals and empty the waste water container, if
necessary.
The waste water container can be removed from and replaced in
the injector as shown in the picture.

7.3 Replacing the heater

Before replacing the heater, disconnect the injector from

the power supply.

The heater system of the IvoBase Injector has been developed in
such a way that users may replace it by themselves, if required
(cleaning, defect).

Dismounting the heater:
To dismount the heater, please
proceed as follows:

Step 1:
Remove the two knurled screws at the rear and remove the hood.