Ivoclar Vivadent IvoBase Injector User Manual

Page 22

Advertising
background image

22

Step 9:
Analyze the measurement. The measuring results must be in the
following temperature range:
– Min.: 90 °C / 194 °F
– Max.: 110 °C / 230 °F
If the measured value is not within this range, please contact your
Ivoclar Vivadent Service Centre.

Step 10:

Press the Stop key to end the test. After that, remove the

flask from the injector and remove the measuring point.

7.7 Service Hint

When the Service Hint is displayed for the first time (Hint 1700),
3 years have passed or 10,000 cycles have been carried out. There-
fore, Ivoclar Vivadent recommends having a maintenance procedure
performed on the injector. The interval until the next appearance of
the Service Hint can be changed in the Settings (see Section 5.4.1).

7.8 Information about transport protection

In order to transport the Ivobase Injector, the clamps must be
brought into a transport position. This function can be activated in
the Settings (see Chapter 5.4 Settings and information).