7.2 Monitoring and maintenance

The time for these maintenance procedures depends on the
frequency of use and the working habits of the users. For that
reason, the recommended times are only approximations.

This furnace has been developed for typical use in dental
laboratories. If the product is used in a production enter-
prise, for industrial applications and for continuous use,

premature ageing of the expendable parts has to be expected.

The expendable parts are as follows:
– Heating muffle
– Insulation material

Expendable parts are not covered by the warranty. Please also
observe the shorter service and maintenance intervals.

If the furnace head is replaced with a furnace head from
another Programat S1, calibration is required.

7.3 Cleaning

The furnace may only be cleaned when it is cool, since

there is a burn hazard. Do not use any cleaning solutions.

Disconnect the furnace from the power source before

cleaning.

The following parts have to be cleaned from time to time:

Item:

Frequency:

Cleaning Material:

Furnace housing (7) and
furnace head

if required

soft, dry cloth

Membrane-sealed
keypad (8)

weekly

soft, dry cloth

Cooling tray (26)

daily

cleaning brush*

Isolation (1)

daily

cleaning brush*

*Never use compressed air!

What:

Measure:

When:

Check all plug-in connections for correct fit

Var. external connections

weekly

Check if the furnace head opens smoothly and without excessive noise

Opening mechanism

monthly

Check if the thermocouple is straight and in the right place

Thermocouple (2)

weekly

Check the insulation for cracks and damages. If the insulation is worn down
it has to be replaced by a certified Ivoclar Vivadent Service Centre. Fine hair-
line cracks on the surface of the insulation are harmless and do not
influence the function of the furnace in a negative fashion.

Insulation (1)

monthly

Check the keypad for visible damage. If the keypad is damaged, it has to be
replaced by a certified Ivoclar Vivadent Service Centre.

Keypad (8)

weekly

Check temperature. Use the temperature checking set to check and adjust
the temperature in the furnace.

Firing chamber

twice a year