K-Patents PR-33-AC User Manual

Page 10

Advertising
background image

4

PR-33-AC instruction manual

Document/Revision No. Rev. 1.02

Effective: June, 2014

It is possible to use a WLAN access point to decrease the number of cables (figure 2.3).

WLAN

Access

point

Ethernet

cable

Figure 2.3

Connecting refractometer(s) via wireless.

The maximum distance of a single Ethernet connection is 100 m (incl.

one joint

adapter/coupler), but if longer distances are required, a fiber link may be used to

extend the range (see figure 2.4). The range may be up to several kilometers with a

suitable fiber connection.

Fiber

Media

converter

Media

converter

Ethernet

cable

Figure 2.4

Using fiber link to connect refractometer(s).

2.3.2 IP settings for the Sanitary OEM Refractometer

All Sanitary OEM Refractometers are shipped with the factory default IP address of

169.254.23.33. This address belongs to the Zeroconf addresses (as defined in IETF

standard RFC 3927) so that it can easily be reached from a stand-alone computer (fig-

ure 2.1), usually without changing the network settings of the computer. This address

will remain in the instrument even after a different IP address has been set. The sensor

answers in the address that is first called up after startup.

Note: If there are more than one Sanitary OEM Refractometers in the same network,

this address cannot be used (section 2.3.4).

The IP address of the instrument can be changed through the instrument homepage

(see chapter 5).